GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A288 - Collimator Lamp
Page no. 1389 Chapter 9
JC-DR-A-288.fm
help you determine in which direction to reposition the lamp.
7. Fully secure the lamp, tightening the allen screws (7) to a torque of 1.1 Nm.
8. Reinstall the right collimator cover with the small screws (8) (small Pozi screwdirver).
Reposition accordingly
3
8
7
8