Chapter 9 Page no. 1392
JC-DR-A-290.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A290 - FOV Button
5. Pull the FOV button away from the Tub Head to remove them from the Tube Head.
6. Follow the FOV button W510 cables (2) up to the top of the Tube Head, cutting the cable ties (3) that
secure the cables to the framework.
7. Fully remove the FOV button.
6-2 Reassembly of the FOV Buttons
1. Route the W510 cables of the new FOV button (1) up to the top of the Tube Head and to the Tilt
Board.
2. Connect the W510 cable to the J4 connector on the Tube Tilt board.
3. Use the cable ties (2) that secure the cables to the framework.
4. Put the new FOV button in place on the back of the Tube Head. Secure the new FOV button on each
2
3
1
2