GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A290 - FOV Button
Page no. 1393 Chapter 9
JC-DR-A-290.fm
side of the Tube Head with allen screws (3)(3 mm allen wrench).
5. Replace the tube head covers; refer to Job Card PHY A044 - Remove/Reinstall Gantry Covers on
page 523.
7 COMPLETION
After changing the FOV button, no calibrations are necessary.