GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A312 - Mouse
Page no. 1421 Chapter 9
JC-DR-A-312.fm
6. Disconnect the mouse cable from the ADS computer.
7. Cut the cable ties that secure the monitor cables and mouse cable to the Control Station framework
(3) and harness (4).
8. Carefully pull the harness (5) from the upper and lower parts of the rotative arm.
9. Extract the mouse cable (6) from the hole in the harness, and from the opening of the monitor sup-
4
3
5
7