GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A315 - International Keyboard
Page no. 1445 Chapter 9
JC-DR-A-315.fm
7 PROCEDURE FOR V1/V2 CONTROL STATIONS
7-1 Determine the OpenBoot Version of the SB150 Workstation
1. From behind the Browser, launch a command window (see Launching a Command Line Window on
page 120).
A command window appears with an ADS prompt.
2. Query the OpenBoot version of the SB150 Workstation, by entering:
prtconf -V
The OpenBoot version of the SB150 Workstation appears.
• If the OpenBoot version is lower than 4.10.6, then you cannot change the keyboard with an Inter-
national Keyboard.
• If the OpenBoot version is equal to or greater than 4.10.6, then shutdown the Control Station
using the shutdown button the Browser, and proceed to section 7-2 to change the keyboard with
an International Keyboard.
WARNING
Before proceeding to section 7-2, shut down correctly the equipment and switch off the
Senographe system hospital supply circuit breaker. Lock it in the off position or mark it
clearly with a warning sign to prevent others from applying power.
7-2 Connect the new keyboard
1. Unplug and remove the X-ray Console.
2. Remove Control Station Top Cover. See Job Card PHY A040 - Remove/Reinstall Control Station
Covers on page 503.
3. Remove Control Station Omega Cover. See Job Card PHY A040 - Remove/Reinstall Control Station
Covers on page 503.
4. Disconnect the original keyboard from the back of the ADS Computer, and remove it from the Control
Station.
43
Reference Function
1 Mouse or Trackball USB Link
2 Keyboard USB Link
3 Not used
4 Bar Code Reader USB Link
1
2