Chapter 9 Page no. 1488
JC-DR-A-319.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A319 - Rotative Arm
7 PROCEDURE FOR SISFLE ARM
7-1 Disassembly of Sisfle Arm
1. Carefully pull on the upper silicone mask (1) to remove it from the rotative arm. Store the upper sili-
cone mask in a clean place.
2. Carefully pull on the lower silicone mask (2) to remove it from the rotative arm support. Store the
lower silicone mask in a clean place.
3. Disconnect the mouse cable from the ADS computer.
4. Cut the cable ties that secure the monitor cables and mouse cable to the Control Station framework
1
2