GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A319 - Rotative Arm
Page no. 1487 Chapter 9
JC-DR-A-319.fm
5. Reinstall the Monitor and Mouse Support with the two screws (7) (allen wrench 3/32'').
6. Insert the harness (8) into the rotative arm.
7. Reinstall the LCD Monitor (refer to Job Card D/R A318 - LCD Monitor on page 1469).
8. Check that the three stop brackets (9) are correctly positioned to avoid collision with the radiation
screen.
9. Reinstall the Omega cover then Top cover (refer to Job Card PHY A040 - Remove/Reinstall Control
Station Covers on page 503).
10. Reinstall and reconnect X-ray Console.
7
8
9