Chapter 9 Page no. 1582
JC-DR-A-407.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A407 - Generator Interface Board 400-PL2
7 COMPLETION
Carry out the following calibrations after you have changed the Generator Interface board:
Calibrations - Checks Job Card
Bias Calibration Job Card CAL A003 - X-Ray Tube Focal Bias Voltage Scale Factor on
page 1655
Heating Current Calibration Job Card CAL A001 - Heater Current Scale Factor on page 1649
kV Scale Factor Job Card CAL A011 - Calibration of kV Scale Factor on page 1667
mAs X-ray Calibration Job Card CAL A012 - X-Ray Tube Heater Current on page 1673
AOP/SNR Check Job Card ELE A038 - AOP Mode and SNR Check on page 707