Job Card CFG A001 - Software Reload
• Follow the Install from scratch instructions given in the booklet supplied with the AWS Application
Software CD-ROM to install the software. Before starting the procedure, take note of the require-
ments for Material needed.
• After completing the installation, Refer to Job Card ELE A052 - Back-up/Restore Procedure on page
771:
- Restore AWS parameters from the system backup on the backup media (either floppy diskette or
CD-R). If this backup media is not available, the procedures given in the Job Card ELE A013 -
AWS Configuration on page 573 must be carried out.
- Back up the system configuration to a floppy diskette or a CD-R.
!
Notice:
If the system backup is not done, no backup for IDC and Senographe parameters will exist.
• For V3 Control Stations that use AWS V53.10.10 or later and V4 Control Stations, install the Embed-
ded Publications, see Job Card ELE A044 - Install User Publications on page 755.
• For Senographe systems that use AWS V43.0 or later (on the SB150) or AWS 53.0 or later (on the
U20) or AWS 54.10 or later (on the Z400), if the Senographe system does not use the default AOP
Classic (AOP A) tables configuration, restore as necessary the AOP configuration (see Job Card
ELE A031 - AOP Configuration on page 693).
6-2 AWS Application Reload or Update
• To reload the Application software only, or if a new version of application software must be installed
on your system, use the AWS Application Software CD-ROM used for the current installation, or that
provided for the update, and follow the Update from CD-ROM instructions given in the CD-ROM
booklet.
• For V3 Control Stations that use AWS V53.10.10 or later and V4 Control Stations, reinstall the
Embedded Publications, see Job Card ELE A044 - Install User Publications on page 755.
• For Senographe systems that use AWS V43.0 or later (on the SB150) or AWS 53.0 or later (on the
U20) or AWS 54.10 or later (on the Z400), if the Senographe system does not use the default AOP
Classic (AOP A) tables configuration, restore as necessary the AOP configuration (see Job Card
ELE A031 - AOP Configuration on page 693).
7 IDC SOFTWARE
7-1 IDC Software Reload or Update from CD-ROM
Software reload from CD-ROM is necessary when the IDC Computer has been replaced, or if a new ver-
sion of the IDC software must be installed on your system. Use the following steps to perform an IDC
Software Reload:
1. Back up the system configuration (including the IDC configuration) to the AWS hard disk. Refer to
Job Card ELE A052 - Back-up/Restore Procedure on page 771.
2. Follow the installation instructions given on the back of the CD-ROM case of the IDC Application
Software CD-ROM used for the current installation or supplied for a software update.
3. Restore the IDC directory data from the current system backup from the AWS hard disk. Refer to Job
Card ELE A052 - Back-up/Restore Procedure on page 771.
4. Reload the detector paramaters according to Load Detector Parameters on page 1103 in Job Card
D/R A141 - Image Receptor/Digital Detector (requires the floppy disk or CD-ROM containing param-
eters for the detector).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals