GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CFG A001 - Software Reload
Page no. 1643 Chapter 9
JC-CFG-A-001.fm
7-2 IDC Data Restore from Backup
If lost or corrupted files make it impossible to run the application after rebooting, restore the IDC directory
data from the current system backup. Refer to Job Card ELE A052 - Back-up/Restore Procedure on
page 771.
8 SENOGRAPHE PARAMETERS
8-1 Senographe Software Reload
• If Senographe parameters have been lost or corrupted, reload them from the current system backup.
Refer to Job Card ELE A052 - Back-up/Restore Procedure on page 771.
• If the backup is not available, it will be necessary to recalibrate the Senographe.