Chapter 9 Page no. 1742
JC-CHK-A-004.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CHK A004 - DICOM Printer Adjustments and Checks
2. Select the FFDM Printer patient from the patient list in the Browser.
3. Click the Review Exam button to open a review exam for the selected FFDM Printer.
A review exam for the FFDM Printer appears (similar to the following).
4. If the display WW and/or WL values have been changed, select the Standard auto-windowing option
to return them to the default values of WW =1500 and WL = 2000.
Auto-windowing
options
WW and WL
values