Page no. 1771 Chapter 11
SC-PM-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Scenario PM A001 - Planned Maintenance Steering Guide
Acquisition IQ:
Clean Tube Head Fan Filter (for sites with
tube head dust reduction kit)
Job Card PM A254 - Tube Head Fan Filter Clean-
ing: p. 1783
620 X X X
Clean Collimator Job Card PM A255 - Collimator and Tube Head
Cleaning: p. 1787
12 45 X X X
If Stereo or 2D Biopsy Optical Localizer
option exists, calibrate the LCD Monitor
Job Card ELE A043 - AWS Monitor Calibration on
page 729
12 40 X X X
If the 21-inch 3 MP Monitor option exists,
check the monitor luminance
Job Card PM A300 - 21-inch 3MP Monitor Lumi-
nance Checks on page 1793
12 60 X X X
IQ Tools Flat Field Test Job Card ELE A032 – IQ Tools Flat Field Test: p.
699
6 30XXXXXX
Calibrate Conversion Factor Job Card CAL A045 - Conversion Factor Mea-
surement: p. 1701
12 8 X X X
Check Image Acquisition and ACR Score Job Card ELE A037 - Image Acquisition and ACR
Score: p. 705
6 8 XXXXXX
Calibrate mAs Non-Linearity (Automatic) Job Card CAL A053 - mAs Non-linearity Calibra-
tion: p. 1725
12 8 X X X
AOP calibration Job Card CAL A051 - AOP Calibration: p. 1723 12 8 X X X
AOP mode and SNR check Job Card ELE A038 - AOP Mode and SNR
Check: p. 707
12 15 X X X
Senographe System PM finalization:
Backup Senographe parameters Job Card ELE A052 - Back-up/Restore Proce-
dure: p. 771
6 10XXXXXX