Chapter 3 Page no. 252
Indicators and Switches.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Central Listing
2-3-1. Generator Command Board 400PL1 Connectors
2-3-2. Generator Command Board 400PL1 Fuses
Connector Function
XJ1 To the XJ1 connector on the Supply Command Board 200PL2 (standby power supply)
XJ2 To the XJ5 connector on the Supply Command Board 200PL2 (250 V)
XJ3 To the CN2 connector on the BT Power Supply 400T1 (power supply ±12 V, +5 V)
XJ4 To the HV Tank (heater, bias)
XJ5 To the XJ1 connector on the kVmA Board (HV Tank control)
XJ6 To XJ1 the connector on the Anode Starter Board (+12 V, control signals)
XJ7 To the XJ1 connector on the Inverter Board (control signals)
XJ8 From X-ray Console Cable (27 V, Rx/Tx signal, PREP, and GRAPHIE signals)
XJ9 To the XJ1 connector on the Generator Interface Board 400PL2 (power supply ±12 V, +5 V)
XJ10 To the XJ2 connector on the Generator Interface Board 400PL2 (control signals)
XJ11 Not Applicable
XJ12 To the Rx button
XJ13 To the XJ11 connector on the Supply Command Board 200PL2 (control signals)
Fuse Specification PN Function
F1
250 V, 2.5 A, Time delay 99194291
+5 V BT supply
F2 -12 V BT supply
F3 +12 V BT Supply