GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Central Listing
Page no. 277 Chapter 3
Indicators and Switches.fm
2-9-3. kVmA Board 700PL1 Switches
2-9-4. kVmA Board 700PL1 Potentiometer
DS3 - 12 V Present Green Normal status: ON
ON when the -12 V from the Generator Command
Board 400PL1 is present.
OFF when there is a problem with the -12 V from
the Generator Command Board 400PL1. In this
case, proceed as follows:
• Check the status of the -12 V on the Generator
Command Board 400PL1 indicated by its green
DS11 LED.
If the DS11 LED is ON, change the cable between
the XJ1 connector on the kVmA Board and the
XJ5 connector on the Generator Command
Board 400PL1.
If the DS11 LED is OFF, check whether the -12 V
returns when you disconnect the cable between
the XJ1 connector on the kVmA Board and the
XJ5 connector on the Generator Command Board
400PL1.
If the problem persists (i.e. DS11 is still OFF),
change the Generator Command Board 400PL1.
Switch Function
SW1 Test 120 mA
SW2 Test 50 mA
Switch Function
P1 kV gain input (1)
P2 kV gain input (2)
P3 Vref B1 adjust
P4 Not Applicable
P5 Measure mA adjust

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals