Chapter 3 Page no. 276
Indicators and Switches.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Central Listing
2-9-2. kVmA Board 700PL1 LEDs
LED Function Color Behavior
DS1 -15 V Present Green Normal status: ON
ON when the -15V from the Generator Command
Board 400PL1 is present.
OFF when there is a problem with the -15V from
the Generator Command Board 400PL1. In this
case, proceed as follows:
• Check the status of the 12 V on the Generator
Command Board 400PL1 indicated by its green
DS9 LED.
If the DS9 LED is ON, change the cable between
the XJ1 connector on the kVmA Board and the
XJ5 connector on the Generator Command Board
400PL1.
• If the problem persists, change the HV Tank.
DS2 +15 V Present Green Normal status: ON
ON when the +15 V from the Generator Command
Board 400PL1 is present.
OFF when there is a problem with the +15 V from
the Generator Command Board 400PL1. In this
case, proceed as follows:
• Check the status of the 12 V on the Generator
Command Board 400PL1 indicated by its green
DS9 LED.
If the DS9 LED is ON, change the cable between
the XJ1 connector on the kVmA Board and the
XJ5 connector on the Generator Command Board
400PL1.
• If the problem persists, change the HV Tank.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals