Chapter 3 Page no. 318
Cables.FM
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Cable Lay-out and Pin-out
1-78. W411 - POS-Bus-2 Tilting Connect
Tilt Board PL401 J2 to Arm Distribution Board PL301 J9:
1-79. W412 - HV Cable
Generator HV tank to X-ray tube J2.
1-80. W500 - Power/Lamp Control Cable
DC-DC Board PL402 J4 to Control Board PL403 J9:
PIN (PL401) Designation Type Level PIN (PL301)
PL401/J9-1 H-CAN Network CAN Diff 5 / 2.5V PL301/J9-1
PL401/J9-2 L-CAN Network CAN Diff 2.5 / 0V PL301/J9-2
PL401/J9-3 GND Ground 0 V PL301/J9-3
PL401/J9-4 H-RT line 1 RT-CAN Diff 5 / 2.5 V PL301/J9-4
PL401/J9-5 L-RT line 1 RT-CAN Diff 2.5 / 0V PL301/J9-5
PL401/J9-6 Stop motion RT 0/+5 V PL301/J9-6
PL401/J9-7 H-RT line 2 RT-CAN Diff 5 / 2.5 V PL301/J9-7
PL401/J9-8 L-RT line 2 RT-CAN Diff 2.5 / 0V PL301/J9-8
PL401/J9-9 Reset line RT 0/+5 V PL301/J9-9
PL401/J9-10 GND Ground 0 V PL301/J9-10
PL401/J9-11 H-RT line 3 RT-CAN Diff 5 / 2.5 V PL301/J9-11
PL401/J9-12 L-RT line 3 RT-CAN Diff 2.5 / 0V PL301/J9-12
Designation Level Pin N°
GND 0 V J4-1
Power_good_N 0 V J4-2
Power_good_P 0/12 V J4-3
Lamp_On Open collector J4-4
GND 0 V J4-5
24 V 24 V J4-6