EasyManuals Logo

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #470 background imageLoading...
Page #470 background image
Chapter 6 Page no. 470
JC-PHY-A-016.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A016 - AC Connection
6-1 AC Supply Check
Determine the nominal voltage and frequency of the available power supply.
6-1-1 Check the Nominal Line Voltage
Consult the hospital electrician to determine the nominal line voltage and frequency. Then use the multi-
meter to measure the open-circuit line voltage to check that it corresponds to the nominal line voltage,
±10%. Allowable nominal line voltages are 200 V, 208 V, 220 V, and 240 V.
6-1-2 Line Frequency
The Senographe operates with line frequencies of 50 Hz or 60 Hz without adjustment.
6-2 Voltage Selection - Mains Distribution Rack
Check the position of the input connection wire on the Mains Distribution Rack transformer terminal
block (reference A in the illustration).
If the tap connection does not correspond to the nominal (not the measured) line voltage you determined
in section 6-1, remove and reconnect the W2 TRANSFOR LINE 1 cable to the correct terminal (200 V,
208 V, 220 V, 230 V, or 240 V). Do not change the position of W1 MAINS LINE 2/N cable, as it must
always be on the 0 V terminal.
A
W2 TRANSFOR LINE1 Cable
W1 MAINS LINE2/N Cable
Label on top of transformer indicates terminal positions

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals