Chapter 6 Page no. 472
JC-PHY-A-016.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A016 - AC Connection
6-4 Configure Fuses F1, F2.
To configure these fuses (see illustration 2 for location), follow the indications on the label located on the
bottom of the Generator cabinet, just below the AC supply switch S1.
WARNING
For safe operation of the Senographe, pay particular attention to the information on this
label.
Determine whether the AC supply consists of two phases (phase/phase), or a single phase
and neutral (phase/neutral).
When the supply is single–phase with neutral (phase/neutral), a metal cylinder (delivered
with the system) must be installed instead of F1.
ILLUSTRATION 1 - LOCATION OF FUSES