Chapter 6 Page no. 488
JC-PHY-A-030.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A030 - Generator and Gantry Disconnection Reconnection
12. At the back of the Power Line Filter Box (13), remove the Earth cable (14) (8 mm open-ended
wrench). The Earth cable originates from the white Gantry AC Supply cable (
labeled 2381711).
13. On the Mains Distribution Board 200 PL1, disconnect the cables PH2 and PH1 from the XJ1, XJ3
connectors (8 mm open ended wrench). The cables PH2 and PH1 originate from the white Gantry
AC Supply cable (
labeled 2381711).
13
14
XJ1
XJ3