GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A040 - Remove/Reinstall Control Station Covers
Page no. 511 Chapter 6
JC-PHY-A-040.fm
7 REMOVAL & REINSTALLATION OF V1 CONTROL STATION COVERS
7-1 Omega Cover (Front Cover).
To remove the Omega cover on V1 Control Stations:
1. Remove the four external screws (2) (3 mm allen
wrench) securing the Omega cover.
2. Release the cover by tilting it towards you.
3. Remove the front cover by pulling.
4. Disconnect the ground connections.
Reverse the procedure to replace the cover.
7-2 Top Cover
CAUTION
Be careful not to damage the fragile upper corner
of the top cover.
To remove the top cover:
1. Remove the omega cover (see above).
2. Unplug and remove the X-ray Console.
3. Remove the three front securing screws (4) (3 mm allen
wrench).
4. Release the cover by tilting it towards you.
5. Remove the top cover by lifting.
Reverse the procedure to replace the cover. Ensure that the Control Station covers have been correctly
fastened and all blanking disks are correctly installed.
Remember to reconnect all ground connections!
1
2
3

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals