Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A042 - Remove/Reinstall Generator Covers
Page no. 517 Chapter 6
JC-PHY-A-042.fm
b. Lift the cover away from the floor so the holes (9) in the bottom end of the cover pass past the
rods (10) on each side.
c. Disconnect the ground wire (11) at the Generator end to avoid risk of creating a short circuit.
d. Fully remove the cover and store it in a safe place.
10
9
11