Chapter 6 Page no. 518
JC-PHY-A-042.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A042 - Remove/Reinstall Generator Covers
6-2 Install Generator Covers
The left and right (thinner covers) must be installed before you install the front and back covers.
!
Notice:
Take care when replacing the Generator Cabinet covers, especially after connecting water pipes
to the Conditioner. Check that the covers do not compress the pipes and cause a blockage.
1. Remove all power from the Senographe.
2. For the left and right covers, do the following:
a. Connect the cover ground wire (1) to the Generator cabinet.
b. Lower the cover towards the floor so the holes (2) in the bottom end of the cover pass on to the
rods (3) on each side.
c. Slide the lip of the cover upwards so it passes side of the Generator cabinet. Fully push the cover
on to the top of the Generator cabinet. Then secure the cover to the Generator cabinet with the