Chapter 7 Page no. 538
SC-ELE-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Scenario ELE A001 - Electrical Installation
8 Configure Generator Job Card ELE A009 - Generator Config-
uration on page 561
9 Set Gantry travel limits Job Card ELE A010 - Set Lift Travel
Limits on page 565
Checks
10 Generator Functional Checks Job Card ELE A012 - X-ray Parameter
Check on page 571
11 Refit Generator covers Job Card PHY A042 - Remove/Reinstall
Generator Covers on page 513
12 Gantry Functional Checks Job Card ELE A015 - Gantry Functional
Checks on page 579
13 Collimation Checks Job Card ELE A028 - Collimation
Checks on page 679
14 If you had remove the Gantry Covers
to get the Gantry into the room then
refit Gantry Covers
Job Card PHY A044 - Remove/Reinstall
Gantry Covers on page 523.
15 Manual Compression Check Job Card ELE A017 - Manual Compres-
sion Check and Calibration on page 583
Configuration
16 AWS configuration Job Card ELE A013 - AWS Configura-
tion on page 573
17 Network connection and configuration Job Card ELE A021 - Networking Con-
nection and Configuration on page 625
18 Optionally change the internal IP
address of the Senographe system
Job Card ELE A050 - Changing
Senographe Internal IP Addresses on
page 757
Optional.
Only required if the
hospital network is
the same as the
Senographe inter-
nal network
19 Declare Network Hosts Job Card ELE A022 - Declare Network
Hosts on page 663
20 Declare Worklist Hosts Job Card ELE A023 - Declare Worklist
Hosts on page 671
21 Declare Performed Procedure Step
Hosts
Job Card ELE A024 - Declare Per-
formed Procedure Step Hosts on page
675
22 Declare Printers Job Card ELE A020 - Declare and Set
Printers (Filming Devices) on page 585
23 Change the AOP configuration Job Card ELE A031 - AOP Configura-
tion on page 693
24 Install Embedded Publications Job Card ELE A044 - Install User Publi-
cations on page 755
Optional configuration
Step Action Job Card Comments Done