Scenario ELE A001 - Electrical Installation
Page no. 539 Chapter 7
SC-ELE-A-001.fm
25 Repeat and Reject Analysis Configu-
ration
Job Card ELE A067 - Repeat and
Reject Analysis Configuration on page
791
26 Installation of CAD Option onAWS Job Card ELE A071 - Installation of
CAD option on AWS on page 793
Hardware options
27 Install Required options
Mass Archiver - LaserCam - Review
Workstation
Refer to supplied
manuals
28 Install Bar Code Scanner if present Job Card ELE A042 - Bar Code Scan-
ner (Option) on page 715
Refit covers
29 Refit Control Station Covers Job Card PHY A040 - Remove/Reinstall
Control Station Covers on page 503
Image Quality Checks
30 If Stereo or 2D Biopsy Optical Local-
izer option exists, calibrate the LCD
Monitor
Job Card ELE A043 - AWS Monitor Cal-
ibration on page 729
31 Flat Field Check Job Card ELE A032 – IQ Tools Flat
Field Test on page 699
32 Image Acquisition & ACR Score Job Card ELE A037 - Image Acquisition
and ACR Score on page 705
33 Acquisition in AOP mode Job Card ELE A038 - AOP Mode and
SNR Check on page 707
Detector Dose check (Germany only)
34 Compliance with Germany for Detec-
tor Dose
Job Card ELE A039 - Compliance with
Germany for Detector Dose on page
711
Only in
Germany
Stereotaxy installation and configuration
35 If Stereotaxy Option is present Refer to Stereotaxy SM (2390552)
Completion
36 Complete AWS System configuration
form
Form LBK A001 - System Configuration
Form on page 797
37 Backup configuration information Job Card ELE A052 - Back-up/Restore
Procedure on page 771
38 Back-up/Restore procedure - Manual
record sheets
Form LBK A003 - Back-up/Restore Pro-
cedure - AWS Parameters Manual
Record Sheets on page 801
39 Clean up room and complete paper-
work
Step Action Job Card Comments Done