Chapter 3 Page no. 62
Product-historyDS.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Product history and compatibility
3-1-11 Nephtys M415 (Z400) - Q3/Q4 2011
During Q3 2011, the following non-software related changes were made to the system:
• Forward production Gantries include a new "Type 3" Lift Screw Assembly with an integrated Motor
Brake.
• The FRU strategy changed for the Generator CPU Board, such that it is now delivered with no
EEPROMs, and seperate EEPROM Kits are now available. For more information, see Empty Gener-
ator CPU Board and EEPROM Kits on page 66.
• The System Identification Label on the Generator cover is slightly different.
• Available IHE 54.11 kit for IHE sites.
• New dust kit (PN: 5394527).
3-1-12 Nephtys M416 (Sunburst) - Q1 2012
• New ADS 54.20 Software (
5261381-8) bringing IHE Pixel Spacing Compatibility, IPM fix and an
updated UPS script