Chapter 7 Page no. 644
JC-ELE-A-021.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Configuration
4. At the Hostname for e1000g0 screen, type the Hostname to be used for the Acquisition Workstation
in the Host name for e1000g0 field. Press <Return> when done, and then press the <F2> key to con-
tinue.
Note:
The Hostname will automatically be used as the DICOM A.E. Title for the workstation. If there is
more than one Senographe system on the hospital site, the names chosen must clearly differenti-
ate between them. For example, by include the unit number or name.
Do not use the following characters in the Hostname:
! $ ” ‘ ’ { } [ ] * @ # ? ^ _ / etc.
ONLY USE ALPHANUMERIC CHARACTERS WITHOUT SPACES.
It is also possible to use the characters - (dash) and . (dot).
Ensure that the Hostname is no more than 12 characters long.
The Hostname must be the same value as the Station Name that is defined in Job Card ELE A013
- AWS Configuration on page 573.
5. At the IP Address for e1000g0 screen, type the external IP address to be used for the Acquisition
Workstation in the IP address for e1000g0 field. Press the <Return> key when done, and then press
the <F2> key to continue.
6. At the Subnet for e1000g0 screen, ensure that Yes is selected for the System part of a subnet option,
Enter the host name which ...
A host name must have at ...
Host name for e1000g0
Host Name for e1000g0
Enter the Internet Protocol (IP) address for this ...
IP addresses contain ...
IP address for e1000g0
IP Address for e1000g0