Chapter 7 Page no. 666
JC-ELE-A-022.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A022 - Declare Network Hosts
3. The following field must be completed if the remote host device is a review workstation:
- Query Retrieve: Select Patient, Study, or No Provider; (refer to the Review Workstation DICOM
3.0 Conformance Statement).
Note:
RWS (U60) is not a Q/R provider (it cannot be queried), but some remote stations such as Rad-
works or AW4.0 (Seno Advantage) are Q/R providers.
4. The following field must be completed if the remote host device is an Archival device:
- Storage Commitment: Set to Yes if the Archival device DICOM 3.0 Conformance Statement lists
this feature; set to No if the Statement does not list the feature.
• If Storage Commitment is set to No; set Message Send to Yes if you want an S to appear in the
Report Status column of the Study or Exam list when a study has been sent to the Archival
device.
• If Storage Commitment is set to Yes; the fields Application Entity Title, Network Address, and Port
Number in the lower part of the window are highlighted. You must complete these fields, even if
the information is the same as that already entered. This is necessary to allow the possibility of
images being sent to one remote host and committed to another.
5. On completion, click the Save button to update and close the window.
Note:
Network Remote Hosts are not updated in the /etc/inet/hosts file, but are found in the file /export/
home/sdc/Prefs/SdCRHosts. For this reason, it is not possible to use ping, telnet, spray, and sim-
ilar commands using the hostname. The IP address must be used instead. For example, to ping
the MR operator console from the workstation, do not use /usr/sbin/ping MR01_OC0. Instead, use
the IP address, e.g., /usr/sbin/ping 192.9.200.1.
Though it is not mandatory to declare hosts in the /etc/inet/hosts file when using the SdC NET pro-
tocol, it is recommend you to do it in any case.
You are returned to the Remote Host Selection window.
6. Click the Done button (on V1/V2 Control Stations) or OK button (on V3/V4 Control Stations) to close
the Remote Host Selection window.
7. For Senographe systems using ADS versions 31.27 and above you must reload the Senographe
system IP Filters so that the newly declared network host has access through the Senographe sys-
tem firewall via the DICOM protocol. To do this, perform the following sub-steps:
a. From behind the Browser, launch a command window (see Launching a Command Line Window
on page 120).
A command window appears.
b. Execute the IPFreload.sh script, by entering:
IPFreload.sh
c. When the Password: prompt appears requesting you to enter the root password, enter:
operator
An End of script message appears. The newly declared network host can now access the Senog-
raphe system.
7-2-1 Declaring a DICOM IC (Independent Console)
A new protocol has been implemented in order to solve some problems with the IC current software
Release 5.5. The problem was that echo, delay, repetition, and inversion times display were being multi-
plied by 1000.
Declare the host as any other Dicom3 host, but select the IC Dicom3.0 protocol.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals