GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A052 - Back-up/Restore Procedure
Page no. 777 Chapter 7
JC-ELE-A-052.fm
11. If the Generator data has been restored, the tube heater current must be recalibrated.
See section 8-3-2.
8-2 Restore Procedure on V3/V4 Control Stations
1. From the Browser, launch the Service Desktop (see Launching the Service Desktop on page 120).
2. From the Service Desktop, click the Utilities button .
3. Select System Tools > Restore: the Restore screen like the following appears.
4. Select the appropriate restore procedure: you can restore separate sub-system data (AWS, IDC, or
Senographe) as required.
5. Select the data source (Backup1 or Backup 2 on the AWS disk or CD-R disk).
6. If restoring from the CD-R disk, insert the CD-R disk containing the backup data into the CD-ROM
drive on the AWS unit.
7. A message indicating successful completion appears at the end of the restore procedure.
8. If the IDC data has been restored, the IDC must be rebooted. See section 8-3-1.
9. If the restore is not completed successfully, check the following:
• That the backup source is functioning correctly (AWS hard disk or CD-R disk).
• That communication is correct between the AWS and the sub-system being restored.
10. If the Generator data has been restored, the tube heater current must be recalibrated.
See section 8-3-2.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals