Chapter 7 Page no. 790
JC-ELE-A-065.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A065 - Trackball Configuration
6. To check that buttons 1 and 6 are correctly set for right click:
With the cursor in the Console window, press the keyboard Control key.
Keep the Control key pressed while you do each of the following checks:
• Press button 1:
if the VT Fonts Menu does not appear, repeat the setup procedure for button 1.
• Release button 1:
if the VT Fonts Menu does not close, repeat the setup procedure for button 1.
• Press button 6:
if the VT Fonts Menu does not appear, repeat the setup procedure for button 6.
• Release button 6:
if the VT Fonts Menu does not close, repeat the setup procedure for button 6.