GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A065 - Trackball Configuration
Page no. 789 Chapter 7
JC-ELE-A-065.fm
5. Press and release button 1 and wait for the trackball to stop beeping.
• Button 6: Right click
1. Press and hold buttons 1 and 6 until the trackball beeps.
2. Press and hold buttons 2 and 5. Wait until the trackball beeps and ensure that you release both
buttons before the beep ends (the beep lasts 2 seconds).
3. Press and release button 6 and wait for the trackball to stop beeping.
4. Press and release button 3 and wait for the trackball to stop beeping.
5. Press and release button 1 and wait for the trackball to stop beeping.
7-3 Trackball Setup Checks
The following procedure is suggested as a check for correct trackball setup:
1. Start up the system as usual.
2. With the Browser window displayed on the AWS screen, press the keyboard Front key on V1/V2/V3
Control Stations, or the keyboard Alt + Tab keys together on V4 Control Stations until the Iconbox
window appears.
3. Double-click the trackball button 4 to open the Console window.
• If the Console window does not open, repeat the setup procedure for button 4.
4. To check that buttons 3 and 4 are correctly set for left click:
With the cursor in the Console window, press the keyboard Control key.
Keep the Control key pressed while you do each of the following checks:
• Press button 3:
if the Main Options Menu does not appear, repeat the setup procedure for button 3.
• Release button 3:
if the Main Options Menu does not close, repeat the setup procedure for button 3.
• Press button 4:
if the Main Options Menu does not appear, repeat the setup procedure for button 4.
• Release button 4:
if the Main Options Menu does not close, repeat the setup procedure for button 4.
5. To check that buttons 2 and 5 are correctly set for middle click:
With the cursor in the Console window, press the keyboard Control key.
Keep the Control key pressed while you do each of the following checks:
• Press button 2:
if the VT Options Menu does not appear, repeat the setup procedure for button 2.
• Release button 2:
if the VT Options Menu does not close, repeat the setup procedure for button 2.
• Press button 5:
if the VT Options Menu does not appear, repeat the setup procedure for button 5.
• Release button 5:
if the VT Options Menu does not close, repeat the setup procedure for button 5.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals