GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 863 Chapter 9
ERR-SUB-A-003.fm
102/047 DEBUG_O
NLY
LIFT INVALID INI-
TIALIZATION
ERROR
There has been an initialization error. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
102/048 DEBUG_O
NLY
LIFT INVALID
FORCE SENSOR
DEFECT ERROR
There has been a force sensor defect. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
102/049 DEBUG_O
NLY
LIFT INVALID
FORCE SENSOR
RANGE ERROR
There has been a force sensor range error. The Lift
node normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
102/050 HALT LIFT HBRIDGE
VOLTAGE ERROR
There has been an H-Bridge voltage error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
102/051 HALT LIFT UNEXPECTED
EVENT ERROR
There has been a software error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
102/052 DEBUG_O
NLY
LIFT INVALID GO
ROTATION FAIL-
URE ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
102/053 DEBUG_O
NLY
LIFT INVALID GO
ANGULATION FAIL-
URE ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
103/001 ADVISORY LCD: A06 Lift end
position is reached
On every “StartMotion” command, the IPMS checks
whether the software stop limit for the specified
direction is not already reached. If the limit is
reached, the "StartMotion" command is ignored and
the following error is returned:
Lift
“UpperLimitAlreadyReached” if the lift upper stop
limit is active
“LowerLimitAlreadyReached” if the lift lower stop
limit is active
Operator error. No service action is required.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card