Chapter 9 Page no. 862
ERR-SUB-A-003.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
102/041 WARNING VALID LINE WAS
ACTIVE ON LIFT
NODE
The valid motion line is active at power-up.
1. Change the Gantry CPU Board (see Job Card D/
R A190 - Gantry CPU Board on page 1131).
2. Perform a Collimator Calibration (see Job Card
CAL A029 - Collimator Format Calibration on
page 1683).
3. Perform a Collimation Check (see Job Card ELE
A028 - Collimation Checks on page 679).
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
102/042 WARNING LIFT WAS
STOPPED WITH-
OUT STOP COM-
MAND
The lift node stopped the motion by itself because
the button/pedal responsible for the motion was
released and no stop order was received before the
timeout.
1. Change the Gantry CPU Board (see Job Card D/
R A190 - Gantry CPU Board on page 1131).
2. Perform a Collimator Calibration (see Job Card
CAL A029 - Collimator Format Calibration on
page 1683).
3. Perform a Collimation Check (see Job Card ELE
A028 - Collimation Checks on page 679).
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
102/043 DEBUG_O
NLY
LIFT INVALID
BRAKE DEFECT
ERROR
There has been a brake malfunction. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
102/044 DEBUG_O
NLY
LIFT INVALID
CLUTCH MAL-
FUNCTION ERROR
There has been a clutch malfunction. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
102/045 HALT LIFT INVALID USER
INTERFACE
ERROR
One of the Gantry footswitch pedals have failed.
Press very slowly or very quickly on each Gantry
footswitch pedals until the 102/045 error is triggered
again to identify which pedal is defective.
Replace the left and/or right faulty Gantry footswitch
pedals.
102/046 DEBUG_O
NLY
LIFT INVALID COL-
LIMATOR BUTTON
ERROR
There is a collimator button error.
Ignore this error. If the problem persists, contact your
OLC.