Chapter 9 Page no. 890
ERR-SUB-A-003.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
122/059 HALT COMPRESSION
INVALID
SOFTWARE_ERRO
R
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
122/060 DEBUG
ONLY
COMPRESSION
INVALID
GO_ROTATION_FA
LIURE
Ignore this error. If the problem persists, contact your
OLC.
122/061 DEBUG
ONLY
COMPRESSION
INVALID
GO_ANGULATION_
FALIURE
Ignore this error. If the problem persists, contact your
OLC.
122/062 DEBUG
ONLY
COMPRESSION
INVALID
UNDERSPEED_ER
ROR
Ignore this error. If the problem persists, contact your
OLC.
122/063 DEBUG
ONLY
COMPRESSION
INVALID
POSITION_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/064 DEBUG
ONLY
COMPRESSION
INVALID
DECELERATION_E
RROR
Ignore this error. If the problem persists, contact your
OLC.
122/065 DEBUG
ONLY
COMPRESSION
INVALID
MOVEMENT_REVE
RSE_DIRECTION
Ignore this error. If the problem persists, contact your
OLC.
122/066 HALT COMPRESSION
INVALID
INTERNAL_SOFTW
ARE_ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
122/067 DEBUG
ONLY
COMPRESSION
INVALID
DSP_TIMEOUT_EL
APSED
Ignore this error. If the problem persists, contact your
OLC.
122/068 DEBUG
ONLY
COMPRESSION
INVALID
TRANSMIT_QUEUE
_FULL
Ignore this error. If the problem persists, contact your
OLC.
122/069 DEBUG
ONLY
COMPRESSION
INVALID
LIFT_DOWN_PEDA
L_PRESS
Occurs when the lift down pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
122/070 DEBUG
ONLY
COMPRESSION
INVALID
LIFT_UP_PEDAL_P
RESS
Occurs when the lift up pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
Code Category Text in Log File or
on LCD