GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 889 Chapter 9
ERR-SUB-A-003.fm
122/047 DEBUG
ONLY
COMPRESSION
INVALID
LIFT_UP_PEDAL_E
RROR
Ignore this error. If the problem persists, contact your
OLC.
122/048 DEBUG
ONLY
COMPRESSION
INVALID
COMPRESSION_PE
DAL_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/049 DEBUG
ONLY
COMPRESSION
INVALID
DECOMPRESSION
_PEDAL_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/050 DEBUG
ONLY
COMPRESSION
INVALID
STOP_MOTION_DS
P_ACTIVE
Ignore this error. If the problem persists, contact your
OLC.
122/051 DEBUG
ONLY
COMPRESSION
INVALID
CLUTCH_MALFUN
CTION
Ignore this error. If the problem persists, contact your
OLC.
122/052 DEBUG
ONLY
COMPRESSION
INVALID
USERINTERFACE_
ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/053 DEBUG
ONLY
COMPRESSION
INVALID
COLLIMATOR_BUT
TON_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/054 HALT COMPRESSION
INVALID
RELAY_ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
122/055 DEBUG
ONLY
COMPRESSION
INVALID
UPPER_OPTICAL_
FORK_REACHED
Ignore this error. If the problem persists, contact your
OLC.
122/056 DEBUG
ONLY
COMPRESSION
INVALID
LOWER_OPTICAL_
FORK_REACHED
Ignore this error. If the problem persists, contact your
OLC.
122/057 DEBUG
ONLY
COMPRESSION
INVALID
INITIALZATION_ER
ROR
Ignore this error. If the problem persists, contact your
OLC.
122/058 HALT COMPRESSION
INVALID
UNCONTROLLED_
MOTION
Ignore this error. If the problem persists, contact your
OLC.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card