GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 899 Chapter 9
ERR-SUB-A-003.fm
132/065 DEBUG
ONLY
TILT INVALID
STOP_ON_VALID_
MOTION_DOWN
ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/066 DEBUG
ONLY
TILT INVALID
VALID_MOTION_AC
TIVE ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/067 DEBUG
ONLY
TILT INVALID
STOP_ON_UIF_RE
LEASE ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/068 DEBUG
ONLY
TILT INVALID
LIFT_DOWN_PEDA
L_PRESS ERROR
Occurs when the lift down pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
132/069 DEBUG
ONLY
TILT INVALID
LIFT_UP_PEDAL_P
RESS ERROR
Occurs when the lift up pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
132/070 DEBUG
ONLY
TILT INVALID
COMPRESSION_PE
DAL_PRESS
ERROR
Occurs when the compression pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
132/071 DEBUG
ONLY
TILT INVALID
DECOMPRESSION
_PEDAL_PRESS
ERROR
Occurs when the decompression pedal is
depressed.
Ignore this error. If the problem persists, contact your
OLC.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card