EasyManua.ls Logo

GE Senographe DS - Page 907

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 907 Chapter 9
ERR-SUB-A-003.fm
152/009 DEBUG
ONLY
Internal 13.3V volt-
age is out of spec
limits
The internal 13.3V voltage is out of the specified lim-
its.
Ignore this error. If the error persists, contact your
OLC.
152/010 DEBUG
ONLY
Internal 5V DC volt-
age is out of spec
limits
The internal 5 V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
152/011 DEBUG
ONLY
Negative unregu-
lated voltage is out of
spec limits
The negative unregulated voltage (coming from
transformation) is out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
152/012 DEBUG
ONLY
Positive unregulated
voltage is out of spec
limits
The positive unregulated voltage (coming from trans-
formation) is out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
152/013 DEBUG
ONLY
Internal -5V DC volt-
age is out of spec
limits
The internal -5V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
152/014 DEBUG
ONLY
Internal 5V DC volt-
age is out of spec
limits
The internal 5 V DC voltage is out of the specified
limits.
Ignore this error. If the error persists, contact your
OLC.
152/015 DEBUG
ONLY
Internal reference
voltage for AD con-
versions is out of
spec limits
The internal reference voltage for A/D conversions is
out of the specified limits.
Ignore this error. If the error persists, contact your
OLC.
152/019 DEBUG
ONLY
Internal 0.6 volts is
bad. Measurement
circuit error
The internal 0.6 V is bad due to a measurement cir-
cuit error.
Ignore this error. If the error persists, contact your
OLC.
152/020 DEBUG
ONLY
Internal EEPROM
error
The PDU MCU has failed. This error is also gener-
ated when EEPROM is empty. This behavior is nor-
mal during the first 5 s after the PDU is first turned
on.
Ignore this error. If the error persists, contact your
OLC.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card