EasyManua.ls Logo

GE Senographe DS - Page 910

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 Page no. 910
ERR-SUB-A-003.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
3-7 Error Codes from 160/yyy: CPU
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card
160/000 HALT INTERNAL SOFT-
WARE ERROR
One of the following out of memory errors occurred:
- functional device creation fails
- configuration interface creation fails
- HAL device creation fails
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/001 HALT INTERNAL SOFT-
WARE ERROR
A Common Object Request Broker Architecture
(CORBA) extraction failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/002 HALT INTERNAL SOFT-
WARE ERROR
An unknown attribute change notification, HAL sig-
nal, or Atlas event was received.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/003 HALT INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to read the configuration file.
1. Reboot the system from the X-ray Console.
2. If the problem persists change the CPU Board
and re-calibrate the Collimator afterwards.
3. If the problem persists, contact your OLC.
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
160/004 HALT INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to get the configuration data.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/005 HALT INTERNAL SOFT-
WARE ERROR
An Atlas event, attribute change, or HAL signal sub-
scription has failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/006 HALT INTERNAL SOFT-
WARE ERROR
The Gantry CPU was unable to access the state
machine file.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/007 HALT INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to get an attribute.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.

Table of Contents