ERR SUB A003 - Gantry Error Messages
Page no. 909 Chapter 9
ERR-SUB-A-003.fm
152/031 DEBUG
ONLY
Spare relay failure The spare relay or its driver is faulty or not present
on the board.
Check the following relays:
Relay-X1 16 A
Relay-K1 K2 8 A
Relay-18A 230 V
152/032 DEBUG
ONLY
Internal AD converter
failure
The internal A/D converter has suffered an error.
Ignore this error. No service action is required.
152/033 HALT Generator cannot be
turned off by PDU
The Generator relay has failed or the cabling and
connections between the Generator and PDU are
not correctly established.
Check the error logs to see whether error 0x5010 is
also reported. If yes, relay is damaged and needs
replacing. If not, the relay is fine, and you must check
cabling and connections.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card