Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 917 Chapter 9
ERR-SUB-A-003.fm
161/050 HALT COMPRESSION
COMMUNICATION
ERROR
The compression node failed to communicate with
the Gantry CPU.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
161/051 HALT LIFT
COMMUNICATION
ERROR
The lift node failed to communicate with the Gantry
CPU.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
162/000 INHIBIT INTERNAL SOFT-
WARE ERROR
The Gantry CPU received an unknown error code
from the lift node.
1. Reboot the system from the X-ray Console.
2. If the problem persists, contact your OLC (prob-
lem linked to Bucky).
162/001 INHIBIT TUBE TILT CAN
AFFECT IQ
The Gantry CPU received an unknown error code
from the tilt node.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
162/003 INHIBIT TILT NODE NOT
STARTED
Possible causes:
The tilt node did not start. At power-up, the tilt node
sends an acknowledgment (which can be different
from the first heartbeat signal) to Gantry CPU. If the
tilt node does not send an acknowledgment, the
Gantry CPU creates this error.
The TP1 connector on the Tilt board is closed.
Proceed as follows:
1. Reboot the system from the X-ray Console and
perform a compression motion before a com-
plete Gantry boot.
2. If the problem persists check the error logs. If this
error occurs in conjunction with error 162/006,
there is a possibility that a jumper exists on the
TP1 connector on the Tilt board. In this case you
must remove the jumper from the TP1 connector
on the Tilt board.
3. If the problem persists, change the Tilt Board
(Firmware or Hardware problem) and re-calibrate
compression force sensor and compression
thickness.
4. If the problem persists, check CAN bus.
D/R A266
page
1365
CAL A027
page
1681
CAL A048
page
1709
CAL A050
page
1721
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card

Table of Contents