Chapter 9 Page no. 916
ERR-SUB-A-003.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
161/040 HALT INTERNAL SOFT-
WARE ERROR
The Gantry CPU received an unknown error code
from the rotation node.
1. Reboot the system from the X-ray Console.
2. If the problem persists check then check the
debug text into the error log,
- If the error code is 20535, replace the rotation
potentiometer.
- If the error code is 20545, reboot the system
and if the problem persists with error code 20545
then replace the rotation board.
- if the error code is different contact your OLC.
D/R A227
page
1239
D/R A221
page
1223
161/041 HALT INTERNAL SOFT-
WARE ERROR
The Gantry CPU received an unknown error code
from the compression node.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
161/043 HALT ROTATION
COMMUNICATION
ERROR
Possible causes:
• The rotation node failed to communicate with the
Gantry CPU.
• The TP2 connector on the Rotation board is
closed.
Proceed as follows:
1. Reboot the system from the X-ray Console.
2. If the problem persists check the error logs. If this
error occurs in conjunction with error 167/008,
there is a possibility that a jumper exists on the
TP2 connector on the Rotation board. In this
case you must remove the jumper from the TP2
connector on the Rotation board.
3. If the problem persists, contact your OLC.
161/044 ADVISORY LCD: A04 Compres-
sion status is
unknown
The compression UIF request was rejected by the
Gantry because the stereotix compression status is
unknown.
Operator error. No service recommended actions.
161/045 ADVISORY LCD: A04 Compres-
sion status is
unknown
The rotation UIF request was rejected by the Gantry
because the stereotix compression status is
unknown.
Operator error. No service recommended actions.
161/046 ADVISORY LCD: A04 Compres-
sion status is
unknown
The lift UIF request was rejected by the Gantry
because the stereotix compression status is
unknown.
Operator error. No service recommended actions.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card