GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A007 - IDC Error Messages
Page no. 939 Chapter 9
ERR-SUB-A-007.fm
64 Check/change IDC-Conditioner water pipes.
Change Conditioner.
Change Detector.
66 Repeat calibrations.
Change IDC.
Error Code Dis-
played
Actions (in recommended order)