Chapter 9 Page no. 984
JC-DIAG-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A001 - AWS Sun-VTS Diagnostics
4. For V1/V2 Control Stations, plug a Standard RS-232 Loopback Adapter in port 1 of Digi Board on the
AWS Computer. For V3 Control Stations, plug a Standard RS-232 Loopback Adapter into Slot #1 of
PCI0 on the AWS Computer.
5. Run the full test as follows:
For V1/V2 Control Stations from the command line, enter:
tip /dev/dty/a001s
For V3 Control Stations from the command line, enter:
tip /dev/term/0
The system answers Connected.
6. Type some alphanumeric characters on the keyboard. If the board is good, they are echoed immedi-
ately on the screen
7. To exit the test, type the command:
<Return>~.
Note: It can be necessary to repeat the sequence several times.
8. Reconnect the cable when the test is complete.
12-2 Full Test with Cable
• Type of test: UNIX command
1. Power up the system and wait for the console login: prompt on the AWS monitor.
2. Login as root as follows:
a. At the console login: prompt, login as root by entering:
root
b. When requested to specify the root password by entering:
operator
3. Remove the Senographe cable from the Senographe serial port.
4. Plug a Standard RS-232 Loopback Adapter in the end of the cable.
5. Run the full test as follows:
For V1/V2 Control Stations from the command line, enter:
tip /dev/dty/a001s
For V3 Control Stations from the command line, enter:
tip /dev/term/0
The system answers Connected.
6. Type some alphanumeric characters on the keyboard. If the cable is good, they are echoed immedi-
ately on the screen
7. To exit the test, type the command:
<Return>~.
Note: It can be necessary to repeat the sequence several times.
8. Reconnect the cable when the test is complete.