Job Card DIAG A003 - AWS File System and SCSI Board Diagnostics
Page no. 991 Chapter 9
JC-DIAG-A-003.fm
b. Re-install the AWS system from scratch (see the AWS Software Installation booklet that comes
with the AWS Software CD-ROM).
4 SCSI BOARD TESTS
Note:
SCSI board tests are only applicable to V1/V2 Control Stations.
• Type: Open boot command
1. Go to the ok prompt (see Go to ok Prompt on page 989).
2. At the command line, enter:
probe-scsi-all
3. Relaunch the AWS application (see Relaunch the AWS Application on page 990).
• Result interpretation:
If no SCSI identifier is found, the SCSI board is faulty.
On a system with a cd-writer, two identifiers are detected; the SCSI controller itself and the CD-writer.
An example of a result showing two SCSI identifiers is given below:
/pci@1f,0/pci@1/pci@1/SUNW,isptwo@4