GE Analytical Instruments ©2016 104 of 270 DLM 68100-09 EN Rev. A
Chapter 4: Installation
lines when the Flush is complete. If not, press Flush All again. (If one or more of the reagent, dilution
water, or check standard lines are not flushed, see Chapter 9, "
Troubleshooting.")
11. When the flush is completed, press the Back button to return to the Menu screen.
12. Press the Back button.
13. When ready, press the Start Analysis button. Press Main to return to the Main screen. A graph of the
CO
2
peaks will be displayed. Each peak appears after its peak height is measured, and the measured
NPOC concentration appears at the top of the screen.
14. When the analysis is complete, the CO
2
peaks will be displayed on the screen. If there are no peaks and
the Mass Response axis is only 1.0 µg full scale, then refer to the Troubleshooting section of this manual.
Otherwise, proceed to the next step.
Step 13: Calibrate the Analyzer
Before placing the Analyzer into normal operation, calibrate the Analyzer in a manner suited to your application.
Consult Chapter 5, "
Calibration and Verification" for details on calibration options.