GE Analytical Instruments ©2016 81 of 270 DLM 68100-09 EN Rev. A
Chapter 4: Installation
Table 3: Analog Outputs (TB1)
Pin Signal Function
1Analog 1 (+)
4-20mA Stream #1
2 GND_ISO (-)
3ANALOG 2 (+)
4-20mA Stream #2
4 GND_ISO (-)
5ANALOG 3 (+)
4-20mA Stream #3
6 GND_ISO (-)
7ANALOG 4 (+)
4-20mA Stream #4
8 GND_ISO (-)
9ANALOG 5 (+)
4-20mA Stream #5
10 GND_ISO (-)
11 ANALOG 6 (+) Stream ID
(11 mA, 12 mA, 13 mA, 14 mA, or 15 mA, as
designated in Table 4:
Stream ID
12 GND_ISO (-)
NOTE:
4-20 mA outputs are isolated from system ground.
4-20 mA outputs are not individually isolated from each other and, therefore, are not intended to be
connected to multiple systems with differences in ground potential.
4-20 mA outputs may be configured by the user to output NPOC, IC, TC, or TOC ppm levels.
Table 4: Stream ID
Active Stream Stream ID
Standby Output matches the standby level set for the first
sequentially-ordered active stream, starting with Stream 1
111 mA
212 mA
313 mA
414 mA
515 mA
Error Output matches the error level set for the first
sequentially-ordered active stream, starting with Stream 1
Warning n/a (Does not change current value.)