GE Analytical Instruments ©2016 186 of 270 DLM 68100-09 EN Rev. A
Chapter 8: Maintenance
7. Remove the syringe from the Analyzer by pulling it down and away from the valve.
The washer should remain in the valve. If the washer comes free of the valve when you remove the
syringe, be sure to capture the washer for re installation. You must place the washer back on the
plunger tip when reinstalling.
Figure 77: Removing the Syringe
8. Pull the plunger all the way out of the syringe barrel.
Figure 78: Pulling Out the Plunger