GE Analytical Instruments ©2016 83 of 270 DLM 68100-09 EN Rev. A
Chapter 4: Installation
NOTE:
Binary outputs are isolated from system ground and from each other.
Binary outputs are implemented by relays which are rated for a maximum voltage of 125 VAC or
110 VDC @ 1 AMP. The maximum switching capacity (resistive load) is 62.5 VA, 33 Watts.
The System Fault Alarm logic is set up so an un-powered instrument indicates an error status. The
System Fault Alarm relay is energized when the Analyzer power is on. Errors cause this relay to de-
energize until the error is cleared.
The Analysis Complete Output relay will energize momentarily (pulse) when the analysis is complete for
every measurement.
Table 7: A l arm Out p u t s (TB4)
Pin Signal Function
1 +24V +24 Volts
2 +24V
3 +24V
4GNDSystem
Ground
5GND
6GND
7 BINOUT1_NO User Assigned
Alarm #1
8BINOUT1_COM
9BINOUT1_NC
10 BINOUT2_NO User Assigned
Alarm #2
11 BINOUT2_COM
12 BINOUT2_NC
NOTE:
Binary outputs are isolated from system ground and from each other.
Binary outputs are implemented by relays which are rated for a maximum voltage of 125 VAC or
110 VDC @ 1 AMP.
+24 volts and system ground are provided for local use only and are not intended for connections more
than 15 feet away from the Analyzer.