Chapter 4: Installation
 
Table 5: Binary Input (TB2)
Pin Signal Function
1 BININ1+
Start/ Stop
2 BININ1-
3 BININ2+
RESERVED
4 BININ2-
5 BININ3+
RESERVED
6 BININ3-
7 BININ4+
RESERVED
8 BININ4-
9 BININ5+
RESERVED
10 BININ5-
11 BININ6+
RESERVED
12 BININ6-
NOTE:
Binary inputs are isolated from system ground and from each other.
The + input should be supplied with a voltage from +11 volts to +24 volts and the - input should be 
controlled by a relay or an open collector transistor. 
Tabl e 6: A larm  Outp uts ( TB3)
Pin Signal Function
1BINOUT3_NO
User Assigned Alarm #32BINOUT3_COM
3BINOUT3_NC
4BINOUT4_NO
User Assigned Alarm #45BINOUT4_COM
6BINOUT4_NC
7BINOUT5_NO
Analysis Complete Output8BINOUT5_COM
9BINOUT5_NC
10 BINOUT6_NO
System Fault Alarm11 BINOUT6_COM
12 BINOUT6_NC