Chapter 4: Installation
Additional Installation Equipment
To install the InnovOx On-Line TOC Analyzer, you will also need to provide the following equipment:
• Mounting hardware to support the Analyzer (See “
Step 3: Select a Location for the Analyzer” on
page 73.)
• Mounting hardware to bolt Mounting Stand to wall or floor (if you ordered a Mounting Stand accessory)
• Insulated wire (22-12 American Wire Gauge - AWG, rated to 300 Volts) for analog and alarm outputs
(optional; see “
Step 6: Install Input/Output, Cables, and Accessories” on page 79.)
• Analog (4-20 mA) recorder (optional)
• Insulated wire (12-12 American Wire Gauge, rated to 300 Volts) for AC power conduit (see “
Step 11: Turn
on Power and Configure Basic Settings” on page 102.)
• An external circuit breaker or switch that disconnects both poles of the supply voltage, rated
appropriately
Step 2: Complete the Identification Records
Complete the Identification Records section on “Identification Records” on page 3 by recording the date of
installation and the Analyzer serial number (found on the side of the Analyzer).
Step 3: Select a Location for the Analyzer
Select a location in a well-ventilated area to avoid inhaling any toxic gases that may be emitted from the
Analyzer's waste. Place the Analyzer in a location away from direct sunlight and exposure to extreme
temperatures.
Mounting Stand Installation
The InnovOx On-Line TOC Analyzer is designed to be mounted on a Mounting Stand or on a wall. For an
illustration of the required clearances for the Mounting Stand accessory, see Figure 2
.
WARNING
Operating at elevated temperatures (greater than 40 °C) prevents proper operation, and operating at low
temperatures (below 10 °C) can cause errors in the measurements or damage to the Analyzer.