GE Analytical Instruments ©2016 148 of 270 DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
can be TOC, TC, IC, or NPOC depending on what mode the analyzer is set up to run when the conversion
factor is used.)
Figure 54: Configuring the Factor Equation
5. Press the Save button to save the equation and return to the Factor screen.
6. Press the Factor Units button and use the alpha/numeric keypad to type the units identifier (up to 10
alpha/numeric characters), such as MG PER L (Figure 55
).
Figure 55: Configuring the Factor Units
7. Press the Save button to save the units and return to the Factor screen.